Controllers/Calibrators
PM600 Pressure Measurement Modules

For 6270A Pressure Controller/Calibrator

Key features:

– Fluke Calibration Quartz Reference Pressure Transducer (Q-RPT) technology provides precise measurement with long term stability
– 0.01% reading measurement uncertainty from 30% to 100% of the modules’ span allows for extremely wide workload coverage
– Onboard barometer included with absolute mode pressure modules, allowing them to be used in both absolute and gauge mode

    Description

    Performance specifications
    The performance specifications describe the complete instrumental uncertainty of the Product. The specifications include all relevant error components (linearity, hysteresis, repeatability, resolution, reference standard measurement uncertainty, 1-year drift, and temperature effects). The specifications are provided at a level of confidence of 95 %, k=2, normally distributed. Precision uncertainty includes linearity, hysteresis, repeatability, resolution, and temperature effects.

     

    PM600 Modules measurement specifications
    Specifications are valid from 15 °C to 35 °C.

    ModelAbsolute Mode Range
    (SI Units)
    Absolute Mode Range
    (Imperial Units)
    Gauge Mode Range 3
    (SI Units)
    Gauge Mode Range
    (Imperial Units)
    BRM600-BA100K470 kPa to 110 kPa10 psi to 16 psi
    PM600-BG15K-15 kPa to 15 kPa-60 inH2O to 60 inH2O
    PM600-G100K0 kPa to 100 kPa0 psi to 15 psi
    PM600-G200K0 kPa to 200 kPa0 psi to 30 psi
    PM600-A100K6 kPa to 100 kPa0.9 psi to 15 psi-94 kPa to 0 kPa-13.8 psi to 0 psi
    PM600-A200K10 kPa to 200 kPa1.5 psi to 30 psi-90 kPa to 100 kPa-13.2 psi to 15 psi
    PM600-A350K10 kPa to 350 kPa1.5 psi to 50 psi-90 kPa to 250 kPa-13.2 psi to 35 psi
    PM600-A700K18 kPa to 700 kPa2.6 psi to 100 psi-82 kPa to 700 kPa-12.1 psi to 100 psi
    PM600-A1.4M0.035 MPa to 1.4 MPa5 psi to 200 psi-0.065 MPa to 1.4 MPa-10 psi to 200 psi
    PM600-A2M0.07 MPa to 2 MPa10 psi to 300 psi-0.03 MPa to 2 MPa-5 psi to 300 psi
    PM600-A3.5M0.07 MPa to 3.5 MPa10 psi to 500 psi-0.03 MPa to 3.5 MPa-5 psi to 500 psi
    PM600-A7MATM 2 to 7 MPaATM 2 to 1000 psi0 MPa to 7 MPa0 psi to 1000 psi
    PM600-A10MATM 2 to 10 MPaATM 2 to 1500 psi0 MPa to 10 MPa0 psi to 1500 psi
    PM600-A14MATM 2 to 14 MPaATM 2 to 2000 psi0 MPa to 14 MPa0 psi to 2000 psi
    PM600-A20MATM 2 to 20 MPaATM 2 to 3000 psi0 MPa to 20 MPa0 psi to 3000 psi

     

    Model1-Year
    Instrumental Uncertainty
    (% of reading or % FS, whichever is greater)
    Precision Uncertainty
    (% of reading or % FS, whichever is greater)
    BRM600-BA100K40.01 % of reading0.008 or 0.0024
    PM600-BG15K0.01 or 0.0030.008 or 0.0024
    PM600-G100K0.01 or 0.0030.008 or 0.0024
    PM600-G200K0.01 or 0.0030.008 or 0.0024
    PM600-A100K0.01 or 0.003 1,30.008 or 0.0024
    PM600-A200K0.01 or 0.003 1,30.008 or 0.0024
    PM600-A350K0.01 or 0.003 10.008 or 0.0024
    PM600-A700K0.01 or 0.003 10.008 or 0.0024
    PM600-A1.4M0.01 or 0.003 10.008 or 0.0024
    PM600-A2M0.01 or 0.003 10.008 or 0.0024
    PM600-A3.5M0.01 or 0.003 10.008 or 0.0024
    PM600-A7M0.01 or 0.003 10.008 or 0.0024
    PM600-A10M0.01 or 0.003 10.008 or 0.0024
    PM600-A14M0.01 or 0.003 10.008 or 0.0024
    PM600-A20M0.01 or 0.003 10.008 or 0.0024

     

     

    Notes

    1. For PM600s absolute mode modules used in absolute mode, root sum square (RSS) with 0.007 % of FS (reduced to k=1 by square root of 3).
    formula
    1. ATM is any atmospheric pressure from 70 kPa to 110 kPa (10 psi to 16 psi).
    2. For absolute ranges used in gauge mode there is an additional uncertainty of ±7 Pa for dynamic barometric compensation. When combined with other uncertainties this changes the threshold uncertainty for the PM600-A100K to 0.008 % Span and for the PM600-A200K to 0.004 % Span.
    3. These modules can be used as a barometric reference module.