    Performance specifications
    The performance specifications describe the complete instrumental uncertainty of the Product. The specifications include all relevant error components (linearity, hysteresis, repeatability, resolution, reference standard measurement uncertainty, 1-year drift, and temperature effects). The specifications are provided at a level of confidence of 95 %, k=2, normally distributed. Precision uncertainty includes linearity, hysteresis, repeatability, resolution, and temperature effects.

    PM500 Modules measurement specifications
    Specifications are valid from 15 °C to 35 °C.

    ModelRange
    (SI Units)
    Range
    (Imperial Units)
    Measure
    Mode2

    (% of reading or % FS, whichever is greater) 
    (% of reading or % FS, whichever is greater)
    PM500-G100K0 kPa to 100 kPa0 psi to 15 psigauge0.01 or 0.0050.007 or 0.0035
    PM500-G200K0 kPa to 200 kPa0 psi to 30 psigauge0.01 or 0.0050.007 or 0.0035
    PM500-G250K0 kPa to 250 kPa0 psi to 36 psigauge0.01 or 0.0050.007 or 0.0035
    PM500-G350K0 kPa to 350 kPa0 psi to 50 psigauge0.01 or 0.0050.007 or 0.0035
    PM500-G400K0 kPa to 400 kPa0 psi to 60 psigauge0.01 or 0.0050.007 or 0.0035
    PM500-G600K0 kPa to 600 kPa0 psi to 90 psigauge0.01 or 0.0050.007 or 0.0035
    PM500-G700K0 kPa to 700 kPa0 psi to 100 psigauge0.01 or 0.0050.007 or 0.0035
    PM500-BG1M-0.1 MPa to 1 MPa-15 psi to 150 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BG1.4M-0.1 MPa to 1.4 MPa-15 psi to 200 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BG2M-0.1 MPa to 2 MPa-15 psi to 300 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BG2.5M-0.1 MPa to 2.5 MPa-15 psi to 400 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BG3.5M-0.1 MPa to 3.5 MPa-15 psi to 500 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BG4M-0.1 MPa to 4 MPa-15 psi to 600 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BG7M-0.1 MPa to 7 MPa-15 psi to 1000 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BG10M-0.1 MPa to 10 MPa-15 psi to 1500 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BG14M-0.1 MPa to 14 MPa-15 psi to 2000 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BG20M-0.1 MPa to 20 MPa-15 psi to 3000 psibi-directional gauge0.01 or 0.0050.007 or 0.0035
    PM500-BA120K360 kPa to 120 kPa8 psi to 17 psiabsolute0.01 % of reading0.005 % of reading
    PM500-A120K30.08 kPa to 120 kPa0.01 psi to 16 psiabsolute0.01 or 0.0050.007 or 0.0035
    PM500-A160K30.08 kPa to 160 kPa0.01 psi to 23 psiabsolute0.01 or 0.0050.007 or 0.0035
    PM500-A200K30.08 kPa to 200 kPa0.01 psi to 30 psiabsolute0.01 or 0.0050.007 or 0.0035
    PM500-A350K0.08 kPa to 350 kPa0.01 psi to 50 psiabsolute0.01 or 0.0050.007 or 0.0035
    PM500-A700K0.08 kPa to 700 kPa0.01 psi to 100 psiabsolute0.01 or 0.0050.007 or 0.0035
    PM500-A1.4M0.035 MPa to 1.4 MPa5 psi to 200 psiabsolute0.01 or 0.0050.007 or 0.0035
    PM500-A2M0.07 MPa to 2 MPa10 psi to 300 psiabsolute0.01 or 0.0050.007 or 0.0035

     

    ModelRange
    (SI Units)
    Range
    (Imperial Units)
    Measure
    Mode2
    (% FS + % of reading)(% FS + % of reading)
    PM500-G2.5K0 kPa to 2.5 kPa0 inH2O to 10 inH2Ogauge0.03 + 0.020.015 + 0.01
    PM500-G7K0 kPa to 7 kPa0 inH2O to 30 inH2Ogauge0.01 + 0.010.005 + 0.005
    PM500-G14K0 kPa to 14 kPa0 inH2O to 50 inH2Ogauge0.01 + 0.010.005 + 0.005
    PM500-G20K0 kPa to 20 kPa0 inH2O to 80 inH2Ogauge0.01 + 0.010.005 + 0.005
    PM500-G35K0 kPa to 35 kPa0 psi to 5 psigauge0.01 + 0.010.005 + 0.005
    PM500-G70K0 kPa to 70 kPa0 psi to 10 psigauge0.01 + 0.010.005 + 0.005
    PM500-NG100K-100 kPa to 0 kPa-15 psi to 0 psinegative gauge0.01 + 0.010.005 + 0.005
    PM500-BG1.4K-1.4 kPa to 1.4 kPa-5 inH2O to 5 inH2Obi-directional gauge0.03 + 0.020.015 + 0.01
    PM500-BG2.5K-2.5 kPa to 2.5 kPa-10 inH2O to 10 inH2Obi-directional gauge0.03 + 0.020.015 + 0.01
    PM500-BG3.5K-3.5 kPa to 3.5 kPa-15 inH2O to 15 inH2Obi-directional gauge0.01 + 0.010.005 + 0.005
    PM500-BG7K-7 kPa to 7 kPa-30 inH2O to 30 inH2Obi-directional gauge0.01 + 0.010.005 + 0.005
    PM500-BG14K-14 kPa to 14 kPa-50 inH2O to 50 inH2Obi-directional gauge0.01 + 0.010.005 + 0.005
    PM500-BG25K-25 kPa to 25 kPa-100 inH2O to 100 inH2Obi-directional gauge0.01 + 0.010.005 + 0.005
    PM500-BG40K-40 kPa to 40 kPa-6 psi to 6 psibi-directional gauge0.01 + 0.010.005 + 0.005
    PM500-BG60K-60 kPa to 60 kPa-9 psi to 9 psibi-directional gauge0.01 + 0.010.005 + 0.005

     

    ModelRange
    (SI Units)
    Range
    (Imperial Units)
    Measure
    Mode2
    % FS% FS
    PM500-BG100K-100 kPa to 100 kPa-15 psi to 15 psibi-directional gauge0.010.005
    PM500-BG200K-100 kPa to 200 kPa-15 psi to 30 psibi-directional gauge0.010.005
    PM500-BG250K-100 kPa to 250 kPa-15 psi to 36 psibi-directional gauge0.010.005
    PM500-BG350K-100 kPa to 350 kPa-15 psi to 50 psibi-directional gauge0.010.005
    PM500-BG400K-100 kPa to 400 kPa-15 psi to 60 psibi-directional gauge0.010.005
    PM500-BG700K-100 kPa to 700 kPa-15 psi to 100 psibi-directional gauge0.010.005

     

    Notes

      1. The 1 Year Instrumental Uncertainty is specified with a zeroing technique in the Operators Manual. If not adhered to the 1 Year Instrumental Uncertainty is:
    1. PM500 gauge or bi-directional mode modules support absolute mode measurement when used with a Barometric Reference Module. Instrumental uncertainty for gauge mode modules used in absolute mode by addition of a barometric reference module is calculated as the uncertainty of the gauge mode module root sum squared with the uncertainty of the barometric reference module. Uncertainty for gauge mode assumes routine zeroing which is default operation mode when used in a chassis.
    2. These modules can be used as barometric reference modules.